In the RTPCR002 kit, the lack of amplification of the N target, despite the amplification of the E target, is due to a mutation in the viral RNA at the probe's target sequence. We recently identified several SARS-CoV-2 sequences with a single mutation that caused the failure of amplification for this target.

This change represents only 2% of the sequences available since January 2020. However, a cluster of samples with this mutation could lead to multiple cases with the same alteration.

To resolve this issue, a supplementary probe has been added for the N target.